Radiographic Equipments
Xray rooms vary in design, depending on their purpose. For example,
a room dedicated to upright chest radiography might not have an xray table
because the patients in this room would be standing for their examinations, not
lying down. A room designed for doing gastrointestinal examinations would be
equipped for both radiography and fluoroscopy. A typical room designed for
general radiography is suitable for many different types of xray examinations. In
a hospital setting, the room will be fairly large, perhaps 18 x 20 feet in
size, with wide doors to accommodate hospital beds and stretchers. Physical
features will include the radiographic table, the xray tube and its support
system, an upright image receptor cabinet against one wall, and a shielded
control booth that contains the control console.
The Xray Tube
The xray tube is the source of radiation. Modern multipurpose
xray tube are dual focus tubes. Their cathode assemblies contain two filaments,
one large and one small. It situated in the focusing cup that directs its
electrons toward the same general area on the target portion of the anode. When
the small filament is activated, its electron are directed to a tiny focal spot
on the target. The small filament and focal spot provide finer image detail
when a relatively small exposure is appropriate, for example, when imaging a
small body part such as a toe or wrist. The large filament provides more
electrons and is aimed at a larger target area.
The combination of large filament and large focal spot is
used when a large exposure is required, such as for radiographs of the lumbar
spine or the abdomen, because the large filament provides more electrons and
the large focal spot can better handle the resulting heat at the anode. The anode
is disk-shaped and rotates during the exposure, distributing the anode heat
over a larger area and increasing the heat capacity of the tube. It is the rotation
of the anode that causes the whirring sound just before and after the exposure.
Xray Tube Housing
The xray tube is located inside a protective barel-shaped
housing. The housing incorporates shielding that absorbs radiation that is not
a part of the useful xray beam. The housing protects and insulates the xray
tube itself while providing a base for attachment that allows radiographer to
manipulate the xray tube and to control the size and shape of the xray beam.
Xray Tube Support
The tube housing may either be attached to a ceiling mount
tube hanger or mounted on a tube stand. Both type of mountings provide support
and mobility for the tube. A tube hanger is suspended from ceiling on a system
of tracks to allow positioning of the tube at locations throughout the room. This
ceiling mount is useful when positioning the tube over a stretcher or when
moving the tube for use in different locations. A tube stand is a vertical
support with a horizontal arm that supports the tube over the radiographic
table. The tube stand rolls along a track that is secured to the floor (and sometimes
also the ceiling or wall), permitting horizontal motion.
A system of electric locks holds the tube support in
position. The control system for all, or most, of these locks is an attachment
on the front of the tube housing. To move the tube in any direction, the
locking device must be released. Moving the tube without first releasing the
lock may damage the lock, making it impossible to secure the tube in position. Do
not attempt to move the tube without first releasing the appropriate lock.
Typical tube motions include the following:
- Longitudinal – along the long axis of the table
- Transverse – across the table, at the right angles to longitudinal
- Vertical – up and down, increasing or decreasing the distance between the tube and the table
- Rotation- allows the entire tube support to turn on its axis, changing the direction in which the tube arm is extended
- Roll (tilt, angle)- permits angulation of the tube along the longitudinal axis and also allows the tube to be aimed at the wall rather than the table.
Detend
A detent is a special mechanism that tends to stop a moving part in a specific location. Detents are built into tube supports to facilitate placement at standard locations. For example, a vertical detent will indicate when the distance from tube to image receptor is 40 inches, a common standard distance. Other detents provide “stop” when the transverse tube position is centered to the table and when the tilt motion is such that the central ray is perpendicular to the table or to the upright image receptor cabinet against the wall.
Collimator
Another attachment to the tube housing is the collimator, a
boxlike device mounted beneath the port, the opening of the housing. Collimators
allow the radiographer to vary the size of the radiation field and to indicate
with a light beam the size, location, and center of the field. There is usually
a centering light that also helps align the image receptor. Controls on the
front of the collimator allow the radiographer to adjust the size of each
dimension of the radiation field. The collimator has a scale that indicates each
dimension of the field at specific source-image distances. A timer controls the
collimator light, turning it off after a certain length of time, usually 30
seconds. This helps to avoid accidental overheating of the unit by prolonged use
of its high intensity light.
While many collimators are manually controlled, some are
equipped with a feature called positive beam limitation (PBL). These collimators
have sensors that detect the size of the image receptor. Some automatically
adjust the radiation field size to the size of the image receptor; others
prevent exposure until the field has been manually adjusted to the size of the
image receptor or smaller. These PBL devices were legally required, for reasons
of radiation safety, to be installed on machines manufactured, moved, sold, or
significantly upgraded during the years between 1970 and 1993. This requirement
no longer exists, but many collimators with PBL features are still in use.
Radiographic Table
The radiographic table is a specialized unit that is more
than just a support for the patient. While the table is usually secured to the
floor, it may be capable of several types of motion: vertical, tilt, and
floating tabletop.
For vertical table motion, a hydraulic motor, activated by a
hand, foot, or knee switch, raises or lowers the height of the table. This allows
lowering of the table so that the patient can sit down on it easily and also
permits the table to rise to a comfortable working height for the radiographer.
Adjustment to exact stretcher height can be made to facilitate patient
transfers. There will be a detent or standard position for routine radiography.
This standard table height corresponds to indicated distances from the xray
tube. Because it is important that standard tube/IR distances be used, it is
necessary to return the table to the detent position after lowering it for
patient access. Not all table are capable of vertical motion.
Tilting Table
A tilting table also uses a hydraulic motor to change
position. In this case, the table turns on a central axis to attain a vertical
position. This allows the patient to be placed in a horizontal or vertical
position or at any angle in between. The table may also tilt in the opposite
direction, allowing the patient’s head to be lowered at least 15 degrees into
the Trendelenburg position. A detent stops the table in the horizontal
position. Tilting is an essential feature of most fluoroscopic tables and may
also be a feature of a radiographic unit.
Special attachment for the tilting table including a
footboard and a shoulder guard to provide safety for the patient when tilting
the table. Pay particular attention to the attachment mechanism so that you
will be able to apply these attachments correctly when needed. Before tilting a
patient on the table, always test the footboard or shoulder guard to be certain
that it is securely attached.
Motor of Tilting Table
The motor that tilts the table is quite powerful and can
overcome the resistance of obstacle placed in the way. Many step stool and
other pieces of movable equipment have been damaged because they were under the
end of the table and out of view when the table motor was activated. Such a
collision can also damage the table motor. Be certain that the spaces under the
head and foot of the table are clear before activating the tilt motor.
Floating Tabletop
A floating tabletop allows the top of the table to move
independently of the remainder of the table for ease in aligning the patient to
the xray tube and the image receptor. This motion may involve a mechanical
release, allowing the radiographer to shift the position of the tabletop, or
the movement may be power-assisted, activated by a small control pad with
directional switches. Power assisted movement is common for fluoroscopic table.
Grids and Buckys
When primary
radiation encounters matter, such as the patient or the xray table, the
resulting interaction produces scatter radiation. Most of the scatter radiation produced
during an exposure originates within the patient. This scatter radiation causes
fog on the radiographic image, a generalized exposure that compromises the
visibility of the anatomic structures. Grids and buckys are devices to prevent
scatter radiation from reaching the image receptor and degrading the image.
Bucky
A bucky is usually located beneath the table surface. It is
a moving grid device that incorporates a tray that holds the Image receptor. The
entire unit can be moved along the length of the table and locked into position
where desired.
Grid
The grid that is incorporated into the bucky device is situated between
the tabletop and the image receptor. It is a plate made of tissue-thin lead
strip, mounted on edge, with radiolucent interspacing material. The strips must
be carefully aligned to the path of the primary xray beam, so precise alignment
of the xray tube is essential. In most radiographic units, the grid moves during the exposure. The purpose of moving the grid is to blur
the image of the thin lead strips so that they are not visible on the
radiograph. When the table has a floating tabletop, the bucky mechanism and
image receptor do not move with the tabletop.
Stationary grids that do not move during the exposure serve
the same purpose as a bucky. A grid may also be incorporated into a device
called a grid cap, which is a grid mounted in a frame that can be attached to
the front of an image receptor for mobile radiography and other special
applications.
Grid or bucky are generally used only for body parts that
measure more than 10 to 12 cm in thickness. (The average adult’s neck or knee
measures 12 cm). when a grid is not needed, the image receptor is placed on the
tabletop.
Upright Image Receptor Units
The upright bucky or the grid cabinet is a device that holds
the image receptor in the upright position for radiography. It is adjustable in
height and may incorporate either a bucky or a stationary grid. When a
stationary grid is included, this device may be referred to as a grid cabinet;
when the grid moves during the exposure, the device is called an upright bucky.
Even when the table tilts to the upright position, it is common to have a
separate upright unit for some examinations such as the cervical spine and the
chest. When the patient is sitting or standing at the upright bucky, the tube
is angled to direct the xray beam toward the image receptor. The distance may
be adjusted to 40 or 72 inches, depending on the requirements of the procedure.
Transformer
Cables from the tube housing connect the xray tube to the
transformer, which provides the high voltage necessary for xray production. Some
transformers look like a large box or cabinet, which may or may not be located
within the xray room. Newer transformer designs are much smaller and may be
incorporated into the control console.
Control Console
The control console, located in the control booth, is the
access point for the radiographer to determine the exposure factors and to
initiate the exposure.
Radiographic control consoles have buttons, switches, dials,
or digital readouts for some or all of the following functions:
- Off/On – controls the power to the control panel
- mA – allows the operator to set the milliamperage, the rate at which the xrays are produced; determines the focal spot size
- kVp – controls the kilovoltage, and thereby the wavelength and penetrating power, of the xray beam.
- Timer – controls the duration of the exposure
- mAs – some units have an mAs control instead of mA and time settings. The mAs ( the product of mA and time) determines the total quantity of radiation produced during an exposure
- Bucky – Activates the motor control of the bucky device so that the grid will move during the exposure
- Automatic exposure controls (AECs) – special settings available on certain units that allows termination of exposure when a certain quantity of radiation has reached the image receptor
- Meters or digital readouts - to indicate the status of the settings
- Prep (ready or rotor ) switch – prepares the tube for exposure and must be continuously activated until exposure is complete
- Exposure switch – initiates the exposure and must be continuously activated until the exposure is complete
- Accessories – other controls may also be present, depending on the equipment and its specific features
Continue reading on Fluoroscopy Xray Equipment
No comments:
Post a Comment